MS-250 Mixed Signal Oscilloscope Option
Width Trigger
Width triggers upon finding a positive- or negative-going pulse width when measured at the
specified voltage level.
On the Trigger dialog, select Width trigger type to display the controls.
1. Choose the Source input line.
2. Choose the line Polarity on which to trigger.
3. Choose the Logic Family that marks the High-Low logic threshold. To enter a custom
threshold, choose Logic Family User Defined and enter the voltage Level.
NOTE: The Logic Family will default to any Logic Setup associated with that line in a
previous digital group setup.
4. Use Width Condition is settings to create an expression describing the triggering pulse
width. This may be:
l Any width Less Than an Upper Value.
l Any width Greater Than a Lower Value.
l Any width In Range or Out Range of values. You may describe the range using
either:
l Limits, an absolute Upper Value and Lower Value.
l Delta, any Nominal width plus or minus a Delta width.
22